Kjeltec 8400 Analyzer Unit User Manual

66

Foss™ Tecator™ Kjeltec™ 8400 Analyzer Unit The ultimate solution for automation in the routine Kjeldahl laboratory. Manufacturer: Foss North America 60063882 Catalog No. Please call Customer Service at 1-800-234-7437 or send an email to. FOSS KJELTEC 8400 KJELDAHL ANALYZER UNIT (AUTO DISTILLATION/ COLORIMETRIC TITRATION) Product Features. This fully automated Kjeldahl analyser is a highly versatile distillation and titration instrument, using the global reference method for protein determination. Ensures optimal throughput and the highest safety standards for use in any laboratory.

View and Download Nikon D3000 user manual online. D3000 Digital Camera pdf manual download. Nikon d3000 manual mode. Free Download Nikon D3000 PDF User Manual, User Guide, Instructions, Nikon D3000 Owner's Manual. This lightweight Nikon D3000 DSLR features a 10.2 MP DX-format CCD sensor plus EXPEED Image Processor assures breathtakingly rich image quality, managing color, contrast, exposure, noise and speed.

Operation

  1. Turn power on.
  2. Turn printer on.
  3. Turn water on.
  4. Go to manual mode.
  5. Flick titrating tube to get bubbles out. Dispense titrant several times to get bubbles out.
  6. Go to steam on. Press arrow key to turn steam on. Allow to warm up for 5 minutes.
  7. Press arrow key to turn steam off.
  8. Go to analysis mode.
  9. Select program.
  10. Go to result. Select blank.
  11. Have tube containing water in place.
  12. Close door.
  13. Run at least two blanks. Make sure the blank value is around 0.10 or less.
  14. Go to result and change to % Nitrogen. (The blank number above should read the last blank value you ran.)
  15. Put a tube with 1 mL of the 1% N solution standard and some water in place. Input 1.0 for the weight.
  16. Close door and let run. Readout should be close to 1% (+ 0.05%).
  17. Proceed with samples. Run the glycine check first. Input the correct weight for each sample. (The result for the glycine check should be close to 18.66 % N.)
  18. If you have blanks within the set run a tube with just water first to flush out the system, then run the blank. (Change result to blank for running the blank then back to % N for the samples. Input the weight for the samples as above.)
  19. Record numbers from printout into nitrogen book and onto lab sheets with 2 decimals.
Kjeltec 8400 Analyzer Unit User Manual

Calculations

  1. %N = 14.01 x N H2SO4 x (mL acid)/(sample wt(g) x 10)
  2. % Crude Protein = %N X 6.25

Foss Kjeltec 8200

Foss kjeltec 8100

Shut Down and Maintenance

Daily

  • Place tube with water in unit. Select manual. Go to steam on. Press arrow key and allow generator to run for about 5 minutes. Press arrow key to turn steam off. Remove tube and dispose of contents.
  • Turn power off.
  • Turn water off.
  • Turn printer off.
  • Clean drip tray. Clean the shield. Rinse and inspect the rubber adapter. Wipe any spillage with a damp cloth. Squirt water into titration vessel and onto pins to clean.
  • Place empty tube in unit and close shield.

Weekly

Wiring DiagramsThese are the Official Service Manuals of the dealerships, written by Toyota specifically for the vehicle(s) listed. Reassembly. They are incredibly comprehensive with detailed diagrams and exploded views. 2014 buick lacrosse repair manual download. This ensures you get the job done right the first time. Installation.

  • Check and clean around reagent tank screw caps to remove crystals. Every one to three months:
  • Clean alkali pump. Put 2 L of warm distilled water in a beaker. Put alkali tube into beaker. Select manual mode. Select add alkali several times. Empty distillation tube as needed. (A rinse using a half a liter of warm water is sufficient.) Replace hose into alkali tank. Flush out residual water and remove entrapped air.
  • Clean receiver solution dispensing system as above only if contamination has occurred.
  • Check volume of alkali. Go to manual mode. Select the pump you want. Measure dispensed liquid. Adjust as necessary.
  • Check volume of receiver solution. Put drain tube into measuring cylinder. Go to manual mode. Select add receiver. Adjust as necessary.
  • Clean splash head. Put 25 mL distilled water and 25 mL acetic acid into distillation tube. Select manual. Go to steam on for 5 to 10 minutes. Turn off steam. Replace tube with tube containing 50 mL distilled water. Distill for 5 minutes. Repeat the distillation three times with fresh water.

Foss Kjeltec

Yearly

Kjeltec 8100

  • Inspect all tubing and connectors.

Kjeltec 8400 Manual